SoloPro instrument-response removal workflow
Set up an RMRESPONSE job, choose the matching device response, track gain handling and verify corrected output units.
What this procedure covers
The supplied SoloPro instructions describe instrument-response removal for SEG-Y, SEG-D and SAC recordings. Response correction requires the matching instrument response and a known processing history. Keep the uncorrected export and use a separate output directory. This is an English adaptation of the supplied 2025 operating note; interface labels and available response options may vary by installed build.
1. Create a project and task
Open SoloPro, create a project and choose its name and save location. Create a processing task, select RMRESPONSE from the available modules and add it to the workflow. Select the input-data directory and a separate output-data directory.
2. Match the device response
Select the response matching the recorded instrument. The source note shows SmartSolo 5Hz, SmartSolo BD 5s and SmartSolo 2Hz device options. Do not apply a listed option to another model simply because the file can be opened. If the required model or calibration is absent, request the matching response information before processing.
3. Account for gain exactly once
Check the original recording script and SoloLite export settings. If acquisition gain was applied and was not removed during export, select Remove Data Gain with the same gain as the recording script. If gain was already removed in SoloLite, or the recording gain was 0 dB, the source procedure uses None at this stage. Keep a written record of the choice.
4. Choose units and filter corners
Select the required output quantity: velocity, acceleration or displacement. Set the four filter-corner frequencies for the instrument response, data sampling and analysis band. The source gives model-dependent examples; these are not universal settings for every survey. Confirm the usable response band and processing objective before choosing the filter.
5. Validate and run the job
Enter a job name, click Add and confirm the task. Select the task and check whether it can execute. Review the log if validation fails. Use RunJob to process a short test selection, then inspect the resulting files and task log before processing the full dataset.
Check the result
Verify the output quantity and units, retained timestamps, sampling interval and component labels. Compare the corrected and uncorrected waveforms for unexpected amplification, clipping or edge effects. Archive the chosen device response, gain handling, filter corners, software build and output settings alongside the corrected data. These checks support reproducibility; they do not substitute for scientific validation of the processing method.
Source and applicability
SoloPro instrument-response removal operating instructions (Chinese source, 2025). English adaptation by SmartSolo Scientific; source UI identifiers are retained.
This tutorial explains the documented workflow; it does not certify compatibility of every later software or hardware revision. Request the matching manual.
Related instruments
IGU-16HR 5 Hz · IGU-16HR-3C 5 Hz · IGU-16HR-3C 2 Hz · IGU-BD3C-5
Continue the workflow
SoloLite project setup and license checklist
Prepare a SoloLite workstation, organize project and raw-data folders, configure a harvesting rack and verify a pilot download.
Read the English tutorialSoloLite raw-data import, QC and seismic export
Reanalyze copied SmartSolo raw data, inspect Harvest QC and prepare a small continuous-receiver or shot-gather export.
Read the English tutorialSmartSolo App deployment scanning and Bluetooth QC
Match receiver points to instrument serial numbers, review deployment records and export field deployment information from the Android app.
Read the English tutorial