SoloLite raw-data import, QC and seismic export
Reanalyze copied SmartSolo raw data, inspect Harvest QC and prepare a small continuous-receiver or shot-gather export.
Prepare your input and metadata
Use an authorized copy of your own SmartSolo recording and keep an unchanged backup. Record the device model and serial number, recording script, sample interval, gain, component orientation and time reference. Keep receiver deployment information and, for active-source work, shot times and spread geometry with the raw data. This tutorial describes the supplied SoloLite V1.0.1.1 interface; check names and options against your installed build.
1. Download or reanalyze
With the documented harvesting workflow, raw data download into the DCCDATA directory when a station is inserted. If raw data were copied manually, preserve the expected folder structure under DCCDATA\Prospect\Project and use Reanalysis to register the copied files in the project database. Select the specific, missing or all-station option that matches the task. The instructions distinguish this from an ordinary DHR download, which does not require the manual-copy reanalysis step.
2. Inspect Harvest QC
Select the relevant stations, right-click and choose Harvest QC. Open the detailed station records and inspect the parameter history over the acquisition period. Compare station identity, timing coverage, GPS information, battery status and other available QC channels with your field notes. Investigate missing intervals or unexpected status changes before exporting the full survey. A completed transfer alone is not proof that the recording meets the survey requirements.
3. Choose the export workflow
Open the seismic-data export dialog. Use Continue Receiver Gather for a selected continuous time interval, or Shot Gather when shot times and the corresponding spread are available. Check the deployment mapping between station serial numbers and receiver line/point numbers. Review the start and end times in UTC.
The cited build lists SEG-D 2.1, SEG-Y 1.0/2.0, SAC, SEG-2 and MiniSEED output. Choose a format supported by the receiving application and verify the options in your build. For SEG-Y 1.0, the instructions limit a trace to 65,535 samples. Set a suitable file/trace duration and component grouping. Record any gain removal or polarity changes.
4. Prepare, review and run
Click Prepare, inspect the prepared-job list and remove unintended jobs. Run a small station/time selection first. Check the job messages, then open the output in your intended viewer and verify station/component labels, start time, sample interval, sample count and waveform continuity. Proceed with the full export once the pilot output is correct.
Keep a reproducible handover
Deliver the export with the raw-data reference, geometry, timing convention, instrument response information and processing history. If response correction will follow, state whether gain has already been removed to avoid applying the same correction twice.
Source and applicability
SmartSolo Lite Software Instruction, V1.0.1.1, slides 26–34 and 49; documented output formats on slides 4–5.
This tutorial explains the documented workflow; it does not certify compatibility of every later software or hardware revision. Request the matching manual.
Related instruments
IGU-16HR 5 Hz · IGU-16HR-3C 5 Hz · IGU-16HR-3C 2 Hz · IGU-BD3C-5
Continue the workflow
SoloLite project setup and license checklist
Prepare a SoloLite workstation, organize project and raw-data folders, configure a harvesting rack and verify a pilot download.
Read the English tutorialSmartSolo App deployment scanning and Bluetooth QC
Match receiver points to instrument serial numbers, review deployment records and export field deployment information from the Android app.
Read the English tutorialSoloPro instrument-response removal workflow
Set up an RMRESPONSE job, choose the matching device response, track gain handling and verify corrected output units.
Read the English tutorial